Abstract
Background Molecular monitoring of chronic myeloid leukemia patients treated with tyrosine kinase inhibitors is essential for therapeutic stratification. Inter-laboratory reproducibility is, therefore, a crucial issue which requires standardization and strict alignment of BCR-ABL1 values to the international scale. An automated cartridge-based assay (Xpert BCR-ABL Monitor™, Cepheid) had been proposed as a robust alternative to non-automated assays. This study aimed to compare inter-laboratory reproducibility of automated and non-automated quantification, the possibility of converting automated results to the international scale, and the potential economic impact of automation.Design and Methods One hundred and eighteen blood samples from chronic myeloid leukemia patients treated with tyrosine kinase inhibitors were prospectively analyzed in two laboratories using both automated and non-automated assays. The economic evaluation involved a micro-costing study and average costs were assessed as a function of sample throughput.Results Automated assays achieved similar inter-laboratory reproducibility to highly standardized non-automated assays and a short delay (≤6 h) between sampling and blood lysis had a positive impact on inter-laboratory reproducibility. Reporting automated BCR-ABL1 ratios on the international scale was possible using a specific conversion factor which may vary with batches. Cost assessment showed that automated assays could be relevant for annual activity levels below 300 since average costs were lower than those of the non-automated assays.Conclusions The Xpert BCR-ABL Monitor™ assay could be appropriately used in a near-patient setting for routine quantification of e13/e14-a2 transcripts, preferably in partnership with a regional reference laboratory. However, its prognostic impact relative to non-automated quantification remains to be tested prospectively within appropriate clinical trials.Introduction
Treatment response monitoring using serial analyses of BCR-ABL1 mRNA levels by reverse transcription (RT) real-time quantitative Polymerase Chain Reaction (RQ-PCR) is essential for chronic myeloid leukemia (CML) patients treated with tyrosine kinase inhibitors (TKI). It is widely accepted that achieving major molecular response (MMR) is associated with favorable progression-free survival,1,2 and recent data showed that molecular response at three months is predictive of MMR achievement by 24 months in patients able to tolerate imatinib 600 mg once daily.3 It has also been proposed that patients who fail to achieve a MMR by 18 months should undergo therapy review4,5 as should those with significant increase in BCR-ABL1 mRNA levels.4,6
Standardized, accurate and reproducible molecular analyses are, therefore, essential for clinicians to refine therapeutic stratification and make clinical decisions. Inter-laboratory reproducibility is, therefore, a crucial issue that requires standardization6 and strict alignment of BCR-ABL1 values on the international scale (IS), as established by the International Randomized Study of Interferon and STI571 (IRIS) laboratories.2,7,8
Within this context, the cartridge-based microfluidic Xpert BCR-ABL Monitor assay developed by Cepheid, that integrates and automates RNA extraction, RT and RQ-PCR for both e13/14-a2 BCR-ABL1 and ABL1 mRNA on the GeneXpert Dx System automated analyzer, might help improve inter-laboratory reproducibility.9 Previous studies showed that this automated BCR-ABL1 fusion detection system did indeed have similar analytical performance, including specificity, sensitivity and intra-laboratory reproducibility, to non-automated TaqMan based RT RQ-PCR assays.9,10 Sensitivity was assessed at 1 K562 cell amongst 100,000, with a linear range of quantitation of 5 logs. Given that neither of these studies fully addressed inter-laboratory reproducibility and alignment on IS, the present study specifically focused on these aspects. A cost analysis was also carried out in order to assess the comparative economic impact of the automated BCR-ABL1 fusion detection system and non-automated assays, with particular emphasis on the impact of annual analytical activity.
Design and Methods
Laboratories, patients, cells and pre-analytical handling
Two Paris hospital laboratories, referred to as A and B, were involved in the study. Both performed over 500 BCR-ABL1 mRNA quantifications/year. The study involved peripheral blood samples that were harvested from 118 TKI (imatinib, dasatinib or nilotinib) treated CML patients. Informed consenting patients expressing b2/e13-a2 or b3/e14-a2 fusion transcripts were prospectively included in the study. Two series of patients were successively recruited from October 2007 to January 2008 (1 series, n = 65) and during May 2009 (2 series, n = 53). This study was approved by the institutional review board of Saint-Louis Hospital, Paris, and registered as 2007–023.
Two 7mL (EDTA,K3) Vacutainer tubes (Beckton Dickinson, France) were sampled and dispatched at room temperature to both laboratories (1 series) or sent to laboratory A (2 series). In the 1 series, blood samples underwent whole blood lysis within 24 h of sampling, but with variable delays between sampling and lysis between the two laboratories. In the 2 series, two 0.2 mL aliquots of each sample were simultaneously subjected to the pre-analytical step of the Xpert BCR-ABL monitor assay. Under these experimental conditions, the delay between sampling and blood lysis was less than 6 h and was strictly identical between aliquots. One aliquot was immediately analyzed in laboratory A using the Genexpert Dx System while the other was immediately frozen at −70ºC and then sent, frozen, to laboratory B for Genexpert Dx analysis.
BCR-ABL1 mRNA quantification
Both laboratories have a GeneXpert Dx System (Cepheid, France) with 4 analytical positions. On receiving the samples, they were first analyzed on the GeneXpert Dx System, starting from 200 μL as according to the manufacturer’s recommendations, using lots 020.01 and 02601 for series 1 and 2, respectively. PCR efficiency provided for each lot was 92%. Samples were conditioned for RNA extraction using TRIzol (Invitrogen, Cergy Pontoise, France), and BCR-ABL1/ABL1 ratios were evaluated according to standard procedures routinely used in each laboratory. For both the automated and non-automated assays, results were expressed as BCR-ABL1/ABL1 ratios and were transformed to decimal logarithmic values for comparison.
It is worth noting that both laboratories actively participated in the “Europe Against Cancer” (EAC) program and followed tightly standardized procedures at all steps of the analytical process.11 These laboratories have also been involved since 2005 in the French national quality control program organized within the RuBIH (French National Network for Innovative Diagnostics in Onco-Hematology) network, which requires bi-annual evaluation of inter-laboratory BCR-ABL1 reproducibility. During the period in which results were generated for the present study (2007–2009), the interlaboratory variability for RuBIH quality controls between laboratories A and B for the 14 positive (0.02% to 50%) samples undergoing non-automated quantification was maximally 2.1-fold, and none of the negative samples tested positive.
Conversion to international scale
As laboratory A is the French reference laboratory for the EUTOS program,12 it has successfully validated an Conversion Factor (CF) for its routinely used non-automated assays (CF=0.48). Ratios obtained with the Xpert BCR-ABL monitor assay in this laboratory were, therefore, compared to IS according to the Bland and Altman comparison procedure, as described by others.8,13,14
Concordance between automated and non-automated quantification
Results were considered to be concordant for MMR IS when both automated and non-automated assays generated BCR-ABL1/ABL1 values of 0.1% or less for a particular sample. Specific CFs for laboratory A for non-automated and automated assays were used (CF at 0.48 and 2.3, respectively). Since laboratory B had not validated its CF at the time of the study, concordance for MMR IS could only be assessed in laboratory A. With respect to concordance for complete molecular response (CMR), this was achieved when BCR-ABL1 transcripts for a given sample were undetectable by both methods. CMR was assessed in both laboratories A and B since CMR status does not depend on IS. Concordance is defined by the following ratio: number of samples attributed to a given molecular status by both methods divided by number of samples attributed to the same molecular status by either method.
Economic evaluation
The economic evaluation involved a micro-costing study conducted from a societal perspective.15,16 Cost assessments were based on direct observation of labor times, equipment and consumables (disposables and reagents) required for quantitative assessments of BCR-ABL1 transcripts. Cost calculations were based on purchase prices of equipment and supplies (0% and 25% price reductions were envisaged) and on cost standards for the technicians, secretary and physician involved. Following standard practice, a 5% discount rate was used to reflect the opportunity costs of investment in equipment. The overhead services costs (electricity, telephone, etc.) were provided by the Financial Services of hospital B. Cost assessments did not include rent, given the significant variability within a city, a region, or a country.
Average costs were estimated as a function of sample throughput, ranging from one sample a year to an annual activity corresponding to full-time employment of a technician involved in non-automated BCR-ABL1 quantification, or to full-capacity use of the GeneXpert Dx System. Two options were evaluated regarding the allocation of equipment to non-automated BCR-ABL1 quantification. While the first option assumed that the equipment was dedicated to BCR-ABL1 quantification, the second one simulated a polyvalent laboratory in which the equipment was increasingly assigned to BCR-ABL1 quantification with the increase in annual activity. Similarly, with the automated assay, an extra analytical position on the GeneXpert Dx System was assumed to be added when required (up to a maximum of 16 since it is possible to multiplex four GeneXpert Dx Systems) because of increases in annual activity.
The possibility of having to repeat BCR-ABL1 quantification due to problems with sampling, RNA quality, or unexpected results was assessed using a survey addressed to the 45 laboratories performing non-automated BCR-ABL1 quantifications within the French RuBIH network. Respondents to the survey (42%) reported repeat rates ranging from 0% to 28%, and the average rate was 10%. Since the repeated steps may concern the whole process (pre-analytical and analytical) or just the analytical phase, the assessment of additional costs due to repeated steps was based on the average 10% repeat rate applied to costs of technician time, equipment, disposables and reagents involved in both pre-analytical and analytical steps. For the Xpert BCR-ABL Monitor assay, the repeat rate was estimated at 5%, due to the small number of steps likely to require repetition in this automated production process.
In keeping with the general economic principle of efficient use of resources, a criterion defined by average cost stability despite variations in annual activity (reflecting, for example, variations in the demand of BCR-ABL1 quantifications addressed to laboratories ) was used as a marker for perennial laboratory activity. Specifically, annual volumes were identified for which a variation of ±50 annual BCR-ABL1 analyses was associated with a ±1% variation in the average cost.
Results
Inter-laboratory reproducibility (1st series)
Assessment of inter-laboratory reproducibility of BCR-ABL1 mRNA quantification using Xpert BCR-ABL Monitor assays or the non-automated assays was based on 65 peripheral blood samples that were sent to the two laboratories and processed within 24 hours, but with variable delays between the two laboratories. In both laboratories, the BCR-ABL1/ABL1 ratios obtained using both automated and non-automated assays were available for 63 out of 65 samples. As shown in Figure 1, there was a high level of inter-laboratory reproducibility for both automated and non-automated assays, with correlation coefficients of 0.92 and 0.97, respectively. Correlation coefficients were, however, higher for the latter method, since there was a significant dispersion of results for lower levels of positivity, below 0.01% (−2 by log10 ratio, Figure 1) using the Xpert BCR-ABL Monitor assay.
The ratios generated by the Xpert BCR-ABL Monitor assay extended approximately 1.5 log lower (down to between 0.001% and 0.0001%) than those resulting from non-automated assays, which were considered to be negative ratios below 0.01% (Figure 1). The explanation for this apparent discrepancy comes from the different methods used for calculation of BCR-ABL1/ABL1 ratios. The Xpert BCR-ABL Monitor assay generates this ratio from ABL1 and BCR-ABL1 Ct values using the delta Ct method, and a difference of up to 20 Ct between ABL1 and BCR-ABL1 can be obtained, providing minimum measurable ratios of 0.0001%. By contrast, non-automated assays generate ratios based on the number of copies measured, with the lowest measurable copy number being 1 plasmid copy amongst the 20000 ABL1 copies/test, giving a theoretical lower reliable limit of sensitivity of 0.005%. As a result, BCR-ABL1 ratios below 0.01% can be generated by the Xpert BCR-ABL Monitor assay as measurable, although less reproducible, values whereas ratios below 0.01% using non-automated assays are seen as discordant positive/negative results (Figure 1). In the present study, 5 of 63 were discordant positive/negative results by the automated assay, compared to 10 of 63 by the non-automated assay. These results suggest that conversion factors to IS are likely to differ for BCR-ABL1 mRNA quantification between automated or non-automated assays.
Impact of pre-analytical delay on inter-laboratory reproducibility (2nd series)
Given the distinct inter-laboratory reproducibility for Xpert BCR-ABL Monitor assay for ratios below 0.1% (−1 in LogRatio, i.e. the clinically critical level defining MMR) in a setting where pre-analytical latency was variable but always less than 24 hours, a 2 series of 53 samples was analyzed in laboratories A and B, under the same experimental conditions but with reduction of the maximal latency between sampling and lysis from 24 to six hours, and with an identical latency in both laboratories, unlike the first series. Figure 2 shows that, under these conditions, inter-laboratory reproducibility remained high down to at least 0.01% (−2 in LogRatio) and correlation coefficients for the 37 positive ratios reached 0.98, as compared to the 0.92 obtained in the case of variable pre-analytical latency (up to 24 h). Linearity of reproducible quantification was extended to 5 logs, compared to 4 logs with the non-automated assay. This suggests that the positive/negative concordance but quantitative dispersion seen with low level positivity below MMR by the automated method in Figure 1 is likely to result from variable pre-analytical delays which can be transformed into quantitative reproducibility down to at least 0.01% (−2 by LogRatio) if pre-analytical delays are reduced.
Reporting GeneXpert® Dx System results on the International Scale
Thirty-one samples from the first series with ratios greater than 0.01% by the GeneXpert Dx System were used for calculation of the IS conversion factor.8,12 BCR-ABL1/ABL1 ratios obtained in laboratory A using non-automated assays were multiplied by the IS conversion factor assessed by the EUTOS procedure12 (CF=0.48) and were then compared to those obtained in the same laboratory with the Xpert BCR-ABL Monitor assay. Figure 3A shows that the ratios resulting from the Xpert BCR-ABL Monitor assay were systematically lower than the IS ratios, indicating a reproducible bias over the entire measurement domain ranging from 50% down to 0.01%. The regression curve was Y = 0.998 X – 0.357 with R = 0.90. This consistent bias allowed calculation of a conversion factor for alignment on the IS of ratios obtained with the Xpert BCR-ABL Monitor assay. Using the Bland Altman method,8 the mean bias between the Log10Ratios relative to the IS was −0.357, corresponding to an CF of 2.3, expressed in fold change. The differences observed were distributed within a 95% confidence interval of ± 3.1, expressed as fold change (Figure 3B). Validation of this conversion factor was performed using an independent series of 19 samples from the 2 series with ratios greater than 0.01%. Figures 3C and 3D show that applying this conversion factor to Xpert BCR-ABL Monitor assay ratios greatly improved their alignment on the IS. The remaining mean bias was assessed at 1.2 fold greater instead of 2.3 fold lower than IS ratios and a 95% confidence interval was calculated as a ± 2.7 fold change (Figure 3D).
Concordance between methods (1st series)
Concordance between automated and non-automated quantification was assessed using the data generated from the first series of samples, for both MMR IS and CMR. In laboratory A, 35 and 39 samples out of 63 were measured at or below 0.1% IS by both or either methods, respectively, giving a 90% concordance rate (Table 1). All 4 discordant samples gave values below 0.1% with the Xpert BCR-ABL Monitor assay and between 0.15 and 0.3% with the non-automated assay. Concordances for CMR reached 62% (8 of 13) and 71% (12 of 17) in laboratories A and B, respectively (Table 1). Discordant results corresponded to samples measured positive either with automated (n=6, median=0.003%, 0.001%–0.006% range) or non-automated quantification (n=4, median=0.008%, 0.004%–0.06% range). These data show comparable sensitivity of the two techniques with positive/negative discrepancies only involving low level positivity, always below MMR.
Economic evaluation
Since both the Xpert BCR-ABL Monitor and standardized non-automated quantification showed similar inter-laboratory reproducibility, the cost-effectiveness analysis was reduced to a cost-minimization analysis based on data collected through direct observation of labor, equipment and supplies involved. The employment of a full-time technician for non-automated BCR-ABL1 quantification led to an estimated maximum of 2,514 assays per year. In comparison, the time necessary for a quantitative assessment of e13/e14-a2 BCR-ABL1 transcripts using Xpert BCR-ABL Monitor assay was estimated at 3.28 hours (including 2.25 h machine time), leading to an estimated maximum of 3,012 assays per year (consistent with a GeneXpert Dx System with 4 analytical positions).
Figure 4A shows that using the GeneXpert Dx System allowed stabilization of average costs at lower levels of activity compared to the non-automated assay. For annual activity levels over 300 or 600, depending on whether the equipment was assumed to be shared or dedicated to the activity, the average costs of automated assay were, however, systematically higher than those of the non-automated assay. Stability of average costs was seen with annual activity levels over 600 for the automated method, over 1,200 for the non-automated method with shared equipment and at least 1,800 with dedicated equipment (Figure 4A). The corresponding average costs are given in Table 2 with the relative contribution to costs from labor, equipment and supplies. Hypotheses regarding price reductions (0% or 25%) on equipment and supplies had an impact on average costs that was less marked for non-automated than for automated assay (Figure 4A and B). Hypotheses on repeat rates (plus or minus 5%), technician time (10% increase due to the possibility of hazards in production), allocation of equipment (in the case of a polyvalent laboratory), service overheads (20% variation in costs) or discount rates used for the calculation of annual values of equipment (0%, 5% or 8%) all had a low impact on the average costs (ranging from 0.4% to 4%; data not shown). A price reduction of 25% on equipment and 60% on cartridges would align the costs of automated and non-automated assay, also at stabilized annual activity levels.
Discussion
Previous published studies showed that the limit of detection using the cartridge-based Xpert BCR-ABL Monitor assay run on the GeneXpert Dx System assessed in K562 cell spiked blood samples and intra-laboratory reproducibility were similar to those reported for non-automated RT-PCR-based assay with approximate sensitivities of 1 K562 cell in 100,000 and a 5 log range of linear quantification.9,10 We, therefore, chose not to repeat these experiments but to assess reproducibility on patient samples, analyzed in duplicate in two EAC standardized laboratories, with variable pre-analytical times which were systematically less than 24 hours.
The prospective study presented here shows that the automated assay allows similar inter-laboratory reproducibility to standardized non-automated assays even in highly harmonized laboratories applying EAC guidelines and that BCR-ABL1/ABL1 ratios can be converted to the international scale by applying an CF. Moreover, it shows that despite an apparently lower inter-laboratory reproducibility for ratios around and below 0.1% for the Xpert BCR-ABL Monitor assay, agreement between methods remained high, once corrected to the CF.
Reducing the latency between sampling and cell lysis to less than six hours improved inter-laboratory reproducibility to levels at least similar, if not superior, to that of highly standardized non-automated assays. Under these experimental conditions, high inter-laboratory reproducibility was also achieved down to 0.01%, which is the required level for accurate measurement of MMR. At these low levels, positive/negative discordances were relatively frequent when using non-automated quantification and there was no evidence based on patient samples with low level of residual disease that automated quantification was less sensitive than non-automated quantification. The 5 log range of linearity for automated quantification (Figure 2) is in line with that previously described.9,10
Reproducible results at these low levels of minimal residual disease were significantly more frequently obtained when pre-analytical steps were reduced to six hours or less, compared to 24 hours. We emphasize that this situation may only apply to patients undergoing treatment with TKI, in which case the samples remain in contact with the TKI during the pre-analytical phase. A similar impact of pre-analytical delay may not apply to patients not being treated with TKI, such as in the stem cell transplantation setting. The impact of pre-analytical delay, however, merits further investigation since, if these results are confirmed within the context of prospective clinical trials, delays of even 24 hours, necessary for transfer to reference laboratories, may reduce inter-laboratory reproducibility, thus favoring near-patient testing.
Although systematically lower than IS values, the Xpert BCR-ABL Monitor ratios showed a consistent bias over the whole measurement range and could, therefore, be converted to IS by applying a 2.3 CF. This 2.3 CF was subsequently applied to an independent series of samples, resulting in a remaining bias of 1.2 fold (95% confidence interval: ± 2.7 fold). This level of corrected performance is consistent with requirements recommended by Branford et al. as a guarantee of reproducible assessment of molecular response status between laboratories.8 In fact, the level of concordance between methods for MMR IS was measured at 90% in laboratory A. This level of concordance between the two methods is close to that reported by Branford et al. for group 1 laboratories.8 Concordances for CMR were comparable between the two laboratories and positive/negative discordant results were only observed among samples with positivity below 0.1%, in line with previously published data.9,10 These data suggest that there is no significant bias in achieving CMR and its impact on therapeutic decisions, such as stopping TKI.
The Xpert BCR-ABL Monitor assay can, therefore, be used to efficiently measure BCR-ABL1 mRNA levels in virtually all peripheral blood samples taken from TKI treated CML patients expressing e13/e14-a2 BCR-ABL1 fusion transcripts and produce inter-laboratory reproducible ratios that can be expressed on IS. Once converted to the IS, the data presented here suggest that comparable sensitivity can be obtained with automated and non-automated standardized BCR-ABL1 quantification at levels down to 0.001%, if pre-analytical delays are reduced to less than six hours. It is, however, worth noting that conversion factors may vary between Xpert BCR-ABL Monitor assay batches, and needs to be re-evaluated for each new lot/batch of assays. This aspect must be evaluated in collaboration with the suppliers.
Given the significant reduction in the complexity of analysis, it is likely that automated quantification will provide more reliable inter-laboratory results between nonstandardized laboratories than the two EAC laboratories which participated in this study. Indeed, comparison of results with a non-standardized laboratory which used different techniques and house-keeping gene generated, as expected, different quantitative results which could not be normalized using the non-automated method, but comparable results using the Xpert BCR-ABL Monitor assay (data not shown).
Economic evaluation was based on an estimated duration (approximately 3.5 h) for the Xpert BCR-ABL Monitor assay, close to previously published estimations9 and substantially lower than that of non-automated assays. We are not aware of other published studies involving economic evaluation of BCR-ABL1 mRNA quantification, but our cost assessments showed that using the Xpert BCR-ABL Monitor assay could be economically relevant for annual activity levels below 300. At these levels, average costs remained lower than those of non-automated quantification, even on shared platforms where equipment costs are optimized. For higher activity levels, automated quantification was associated, in the absence of significant price reductions on cartridges, with higher average costs than those of non-automated assay, even in the less realistic scenario where the equipment was entirely dedicated to BCR-ABL1 quantification.
Annual activity is likely to increase as the number of patients benefitting from TKI maintenance increases. From an economic perspective, the Xpert BCR-ABL Monitor assay is not optimal for diagnostic screening in order to exclude BCR-ABL1, since it will not detect variant fusion transcripts. This option would require that all negative results (about 90% in a diagnostic setting) be confirmed using non-automated assays in order to detect the approximately 1% of patients with BCR-ABL1 fusion transcripts other than e13-a2/e14-a2. Providing rapid identification of e13a2/e14a2 transcripts using the Xpert BCR-ABL Monitor assay would then be obtained at much higher overall unit cost and could lead to longer delays in providing results. If the automated assay is to be used in this setting, such use should be restricted to cases with a very high index of suspicion.
Overall, the present study showed that the Xpert BCR-ABL Monitor assay could be appropriately used to provide reproducible results for CML patients treated with TKI in a near-patient testing setting. This would allow shorter delays between sampling and blood lysis, with a positive impact on inter-laboratory reproducibility, more rapid (same day) transmission of results, and lower average costs for laboratories with annual activity levels below 300 follow-up samples. Rapid, near-patient follow up of CML patients with classical BCR-ABL1 transcripts treated with TKI could be carried out in local laboratories using the Xpert BCR-ABL Monitor assay, but should be undertaken within a partnership with a reference center performing more specialized molecular analyses, including TKI domain mutations detection, quantification of BCR-ABL1 fusion transcripts other than e13-a2/e14-a2, and follow up of patients following stem cell transplantation. Obviously, automated BCR-ABL1 quantification needs to be undertaken in respect of state-of-the-art management of quality control, training, interpretation of results and record keeping.
Approximately 700 individuals with CML are diagnosed in France every year;17 not all are treated within prospective clinical trials. Clinical guidelines recommend quarterly follow-up testing for patients treated with TKI.4,5 With approximately 20,000 BCR-ABL1 analyses performed in France in 2009 (population 66m), reproducibility and costs are crucial issues. Therefore, therapeutic stratifying cut-off values must be reproducibly generated for patients treated with TKI, whether within a clinical trial or not. While it is likely that patients in trials will undergo molecular monitoring in reference centers, the data presented here, applicable to the vast majority of CML patients, could encourage rational provision of health resources and the establishment of appropriate partnerships between local and reference molecular diagnostic laboratories. This would optimize molecular monitoring and use of TKI.
Acknowledgments
The authors would like to thank all members of the RuBIH (Innovative diagnostics in Onco-Hematology) network for helpful discussions, the Assistance Publique-Hôpitaux de Paris (AP-HP) “Plateforme de Génétique Somatique” and the Saint-Louis tissue resource laboratory for sample preparation.
Footnotes
- Authorship and Disclosures The information provided by the authors about contributions from persons listed as authors and in acknowledgments is available with the full text of this paper at www.haematologica.org.
- Financial and other disclosures provided by the authors using the ICMJE (www.icmje.org) Uniform Format for Disclosure of Competing Interests are also available at www.haematologica.org.
- Funding: this work was supported by the French Direction of Hospitalisation and Healthcare Organisation (DHOS) and by the French National Institute of Cancer (INCA) “Aid for costly technological Innovation” (STIC 2006).
- Received October 5, 2010.
- Revision received January 14, 2011.
- Accepted February 8, 2011.
References
- Druker B, Guilhot F, O’Brien S, Gathmann I, Kantarjian H, Gattermann N. Five-year follow-up of patients receiving imatinib for chronic myeloid leukemia. N Engl J Med. 2006; 355(23):2408-17. PubMedhttps://doi.org/10.1056/NEJMoa062867Google Scholar
- Hughes T, Kaeda J, Branford S, Rudzki Z, Hochhaus A, Hensley M. Frequency of major molecular responses to imatinib or interferon alfa plus cytarabine in newly diagnosed chronic myeloid leukemia. N Engl J Med. 2003; 349(15):1423-32. PubMedhttps://doi.org/10.1056/NEJMoa030513Google Scholar
- Hughes T, Branford S, White D, Reynolds J, Koelmeyer R, Seymour J. Impact of early dose intensity on cytogenetic and molecular responses in chronic phase CML patients receiving 600 mg/day of imatinib as initial therapy. Blood. 2008; 112(10):3965-73. PubMedhttps://doi.org/10.1182/blood-2008-06-161737Google Scholar
- Baccarani M, Saglio G, Goldman J, Hochhaus A, Simonsson B, Appelbaum F. Evolving concepts in the management of chronic myeloid leukemia: Recommendations from an expert panel on behalf of the European LeukemiaNet. Blood. 2006; 108(6):1809-20. PubMedhttps://doi.org/10.1182/blood-2006-02-005686Google Scholar
- Baccarani M, Cortes J, Pane F, Niederwieser D, Saglio G, Apperley J. Chronic Myeloid Leukemia: an update of concepts and management recommendations of European Leukemia Net. J Clin Oncol. 2009; 27(35):6041-51. PubMedhttps://doi.org/10.1200/JCO.2009.25.0779Google Scholar
- Hughes T, Deininger M, Hochhaus A, Branford S, Radich J, Kaeda J. Monitoring CML patients responding to treatment with tyrosine kinase inhibitors: review and recommendations for harmonizing current methodology for detecting BCR-ABL transcripts and kinase domain mutations and for expressing results. Blood. 2006; 108(6):28-37. PubMedhttps://doi.org/10.1182/blood-2006-01-0092Google Scholar
- Branford S, Cross N, Hochhaus A, Radich J, Saglio G, Kaeda J. Rationale for the recommendations for harmonizing current methodology for detecting BCR-ABL transcripts in patients with chronic myeloid leukaemia. Leukemia. 2006; 20(11):1925-30. PubMedhttps://doi.org/10.1038/sj.leu.2404388Google Scholar
- Branford S, Fletcher L, Cross N, Müller M, Hochhaus A, Kim D. Desirable performance characteristics for BCR-ABL measurement on an international reporting scale to allow consistent interpretation of individual patient response and comparison of response rates between clinical trials. Blood. 2008; 112(8):3330-8. PubMedhttps://doi.org/10.1182/blood-2008-04-150680Google Scholar
- Winn-Deen E, Helton B, Van Atta R, Wong W, Peralta J, Wang J. Development of an integrated assay for detection of BCR-ABL RNA. Clin Chem. 2007; 53(9):1593-600. PubMedhttps://doi.org/10.1373/clinchem.2007.085472Google Scholar
- Jobbagy Z, van Atta R, Murphy K, Eshleman J, Gocke C. Evaluation of the Cepheid GeneXpert BCR-ABL assay. J Mol Diagn. 2007; 9(2):220-7. PubMedhttps://doi.org/10.2353/jmoldx.2007.060112Google Scholar
- Gabert J, Beillard E, van der Velden VH, Bi W, Grimwade D, Pallisgaard N. Standardization and quality control studies of ‘real-time’ quantitative reverse transcriptase polymerase chain reaction of fusion gene transcripts for residual disease detection in leukemia - a Europe Against Cancer program. Leukemia. 2003; 17(12):2318-57. PubMedhttps://doi.org/10.1038/sj.leu.2403135Google Scholar
- Müller M, Cross N, Erben P, Schenk T, Hanfstein B, Ernst T. Harmonization of molecular monitoring of CML therapy in Europe. Leukemia. 2009; 23(11):1957-63. PubMedhttps://doi.org/10.1038/leu.2009.168Google Scholar
- Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986; 1(8476):307-10. PubMedhttps://doi.org/10.1016/S0140-6736(86)90837-8Google Scholar
- Bland JM, Altman DG. Measuring agreement in method comparison studies. Stat Methods Med Res. 1999; 8(2):135-60. PubMedhttps://doi.org/10.1177/096228029900800204Google Scholar
- Drummond M, O’Brien B, Stoddart G, Torrance G. Methods for the economic evaluation of health care programmes. Oxford University Press: Oxford, England; 1997. Google Scholar
- Luce B, Manning W, Siegel J, Lipscomb J. Estimating costs in cost-effectiveness analysis. Oxford University Press: New York; 1996. Google Scholar
- Bossard N, Velten M, Memontet L, Belot A, Maarouf N, Bouvier A. Survival of cancer patients in France: a population-based study from the Association of the French Cancer Registries (FRANCIM). Eur J Cancer. 2007; 43(1):149-60. PubMedhttps://doi.org/10.1016/j.ejca.2006.07.021Google Scholar