Anti-doping control of erythropoietin (Epo) relies on the differentiation by isoelectric profile of natural endogenous hormone from the recombinant hormone used for doping. The first and second generations of recombinant Epo were detectable in urine.1,2 The third generation, Continuous Erythropoietin Receptor Activator (CERA), was obtained by linking a methoxy polyethylene glycol to epoetin β, a first generation rHuEPO, resulting in significantly greater stability in blood. CERA, approved in Europe in July 2007, was expected to be quickly misused in sport. Several questions were thus raised about its detectability. We demonstrate here that detection in blood poses no problem and that partial excretion in urine occurs after some physical exercise. Our analytical results from the Tour de France 2008 anti-doping control, which tested both urine and blood samples, serve as an illustration.
Staining revealed that the isoelectric pattern of pure CERA was composed of seven bands with apparent isoelectric points in the pI 4.8–5.2 range (Figure 1). This was surprising since CERA that results from pegylation of epoetin β via the amino group of lysine residues was expected to be more acidic than epoetin β. One explanation could be a shielding effect of the PEG shell on some surface charges of epoetin. In any case, the isoelectric pattern clearly differentiated CERA from natural Epo.
However, analysis of Epo isoelectric patterns in urine or plasma as previously described2,3 required two monoclonal anti-human Epo antibodies. Briefly, 18 mL of urine or 1 mL of plasma were first submitted to a preparation step including immunoaffinity chromatography using clone 9C21D11 and ultrafiltration at a 30-kDa molecular weight cut-off. The retentates from ultrafiltration were then submitted to isoelectric focusing and Epo was detected by double-blotting, using clone AE7A5.4,5 The final result was a chemiluminescent image of the isoelectric patterns of Epo.
Due to the polymer chain in the molecule, the CERA structure might not be recognized by the antibodies used in this method. In fact, use of AE7A5 only (double-blotting only) and both AE7A5 and 9C21D11 (immunochromatography and double-blotting) provided patterns identical to that obtained without antibody (stained gel) (Figure 1).Thus, although the amino terminus of the polypeptidic chain is one of three major pegylation sites, the epitope of AE7A5, which is within the first 26 amino acids, was still accessible to this antibody and the epitope of the 9C21D11 antibody (not identified) was also not affected by pegylation. The sensitivity of the method was determined to be less than 40 pg applied onto the IEF gel (corresponding to initial 50 pg/mL in 1 mL serum and 3 pg/mL in 18 mL urine).
These results proved that, technically, the usual method for Epo analysis could detect CERA. However, due to its hydrodynamic volume that is critical for glomerular filtration, CERA detection in urine, the main biological medium for anti-doping control, seemed unlikely.
Anti-doping control of recombinant Epo during the Tour de France 2008 was performed as usual from urine. CERA was clearly detected in two urine samples (both from the same athlete) reported as positive. Other samples from this same athlete and 3 others gave rise to faint CERA images and were considered highly suspicious. Blood samples taken before the race and mid-race were retrospectively analyzed. All eight plasma samples ( two per athlete) from the 4 athletes having positive or suspicious urine samples gave rise to very clear CERA mages. This confirmed that only a small part of CERA i was excreted from blood into urine. However, strenuous exercise enhances protein excretion in urine by increasing glomerular permeability and decreasing tubular reabsorption.6–8 CERA excretion in urine is affected by exercise as well. This is illustrated by the case of the athlete reported positive from two urine samples. Two positive blood samples were taken 11 days apart while four urine samples were tested for Epo during this period (Figure 2). Sample 1 was totally devoid of Epo (natural or CERA) and sample 3 revealed only traces of drug. Both samples presented physiological protein levels of 36 and 60 mg/L, respectively (Coomassie Plus Protein Assay Reagent, Pierce). The presence of CERA was clearly established in samples 2 and 4, which presented high total protein levels (429 and 227 mg/L, respectively).
In conclusion, the IEF test for anti-doping control of Epo is perfectly able to detect CERA in blood and urine, from a technical point of view. However, the poor drug excretion in urine must be taken into account for anti-doping control purposes and tests may have to be performed in blood.
References
- Lasne F, de Ceaurriz J. Recombinant erythropoietin in urine. Nature. 2000; 405:635. Google Scholar
- Lasne F, Martin L, Crepin N, de Ceaurriz J. Detection of isoelectric profiles of erythropoietin in urine: differentiation of natural and administered recombinant hormones. Anal Biochem. 2002; 311:119-26. Google Scholar
- Lasne F, Martin L, Martin JA, de Ceaurriz J. Isoelectric profiles of human erythropoietin are different in serum and urine. Int J Biol Macromol. 2007; 41:354-7. Google Scholar
- Lasne F. Double-blotting: a solution to the problem of non-specific binding of secondary antibodies in immunoblotting procedures. J Immunol Methods. 2001; 253:125-31. Google Scholar
- Lasne F. Double-blotting: a solution to the problem of nonspecific binding of secondary antibodies in immunoblotting procedures. J Immunol Methods. 2003; 276:223-6. Google Scholar
- Kachadorian WA, Johnson RE. Renal responses to various rates of exercise. J Appl Physiol. 1970; 28:748-52. Google Scholar
- Poortmans JR, Labilloy D. The influence of work intensity on postexercise proteinuria. Eur J Appl Physiol Occup Physiol. 1988; 57:260-3. Google Scholar
- Poortmans JR, Brauman H, Staroukine M, Verniory A, Decaestecker C, Leclercq R. Indirect evidence of glomerular/tubular mixed-type postexercise proteinuria in healthy humans. Am J Physiol. 1988; 254:F277-83. Google Scholar